Use automatic standard addition – Metrohm 757 VA Computrace User Manual

Page 158

Advertising
background image

7 How to ...?

757 VA Computrace – Software

148

13. Start the measurement by clicking the

icon in the

MAIN

WINDOW

or the

button in the

MONITORING

window.

14. Enter the

Sample ID

(used as part of the determination file

name) in the

PLACE SAMPLE

window and click

<OK>

.

15. Each time a solution exchange is required in the

BATCH SO-

LUTION EXCHANGE

window, replace the measuring solution

by the next spiked sample solution.

Use automatic standard addition

In the automatic standard addition, a known amount of the ana-
lyte is added once or several times to the sample using 665 or
765 Dosimats. Proceed as follows:
1. Install Dosimats to the 757 VA Computrace Stand (see How to

install Dosimats).

2. Click on

or

MAIN WINDOW

/ Mode / Determination

.

3. Click on

or

MAIN WINDOW

/ Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

4. Load the desired method into the

WORKING METHOD SPECIFI-

CATIONS

window (see How to load a method).

5. Select

Standard addition

in the

Calibration

field of the

WORKING

METHOD SPECIFICATIONS

window.

6. Select

Automatic

in the

Addition

field of the

WORKING METHOD

SPECIFICATIONS

window.

7. Click

to open the

DOSIMATS

window.

8. Check the Dosimats which are should be used for standard

addition in the

Dosimat used

field (see Dosimats, section 5.2).

9. Close the

DOSIMATS

window by clicking

<OK>

.

10. Click

to open the

EDIT WORKING METHOD PA-

RAMETERS

window.

11. Select the

Determination

tab (see section 5.2) and enter

Sam-

ple identifier

,

Sample amount

,

Cell volume

, and the number of

standard additions in the

No. of additions

field.

12. Select the

Substances

tab (see section 5.2) and make sure

that for each substance entered in the table the number of the
single or mixed standard addition solution, its concentration
and its volume is defined. The

No.

of the standard solution

must be identical to the number of the Dosimat used for
automatic addition of this solution.