8 troubleshooting, 1 general procedure for error messages, 2 connection problems – Metrohm 757 VA Computrace User Manual

Page 171: Error message "could not start the embedded system, Error message "cannot reach the hardware

Advertising
background image

8.1 General procedure for error messages

757 VA Computrace – Software

161

8 Troubleshooting

8.1

General procedure for error messages

Error messages and warnings are displayed in the

CT757

window.

Read the information about the possible causes and the proce-
dure for their rectification and click the

<OK>

button.

8.2 Connection

problems

Error message "Could not start the embedded system"

If this error message appears after starting the VA Computrace
program, the VA Computrace Interface is not running properly.
Proceed as follows:
1. Close the VA Computrace programProgramm.
2. Remove the cable to the mains adapter from the VA Com-

putrace Interface.

3. Reconnect the mains adapter to the VA Computrace Inter-

face.

4. Restart the VA Computrace program.

Error message "Cannot reach the hardware"

If this error message appears after starting the VA Computrace
program, the 757 VA Computrace Stand is not switched on. Pro-
ceed as follows:
1. Switch off the 757 VA Computrace Stand.
2. Wait at least 2 sec.
3. Switch on the 757 VA Computrace Stand.
4. Click the

<Retry>

button in the

CT757

window.