Performing exploratory measurements – Metrohm 757 VA Computrace User Manual
Page 69

4.2 Exploratory specification window
757 VA Computrace – Software
59
Performing exploratory measurements
Measurements in the exploratory mode can be performed using
the following icons (in the
MAIN WINDOW
) or buttons (in the
EX-
PLORATORY SPECIFICATION
window):
Start measurement
The operation sequence (see section 3.4) de-
fined in the
EXPLORATORY SPECIFICATION
win-
dow is started. Each step of the operation se-
quence is listed in the first line of the status win-
dow beside the
<Start>
button.
For the running voltage sweep there is a live dis-
play in the
EXPLORATORY CURVES
window with
automatic scaling of the axes. Manual rescaling
can be done by pressing the <F4> button.
The red light at the left side of the
<Stop>
button
indicates a current overload. In this case, stop
the measurement and change the measurement
parameters.
In the second line of the status window com-
ments or error messages concerning the running
measurement are displayed.
A running measurement can be stopped, inter-
rupted and continued. Each step in the operation
sequence can be abbreviated by clicking the
<Next>
button.
Stop measurement
Stop running measurement immediately.
Hold measurement
Interrupt running measurement.
Continue measurement
Continue an interrupted measurement.
Next step
Go to next step in operation sequence.