3 General settings for exploratory and determination mode

757 VA Computrace – Software

30

RDE/SSE

RDE stands for Rotating Disk Electrode and is used for direct
and stripping determinations with

Solid State Electrodes (SSE).

Stirrer/RDE (rpm)

[ 0...3000 rpm ; 2000 rpm ]

Revolutions per minute of the rotating disk electrode. The
stirring of the RDE remains active during all preparation pro-
cedure steps until the start of sweep.

t

Surface

Notes:
• The RDE can be used for all measurement modes.
• For the 757 VA Computrace Stand, a drive shaft with different

electrode tips is available as an option (see Hardware Man-
ual
).

• For installation and maintenance of the RDE, see Hardware

Manual.