Double peak – Metrohm 757 VA Computrace User Manual

Page 181

Advertising
background image

8.6 Voltammetric problems

757 VA Computrace – Software

171

With

RDE/SSE:

1. The background current is too high: repolish the electrode.
2. The

Deposition time

is too long: reduce the time.

3. Is the

Deposition potential

correct?

Double peak

With

all types of electrodes:

1. Organic components interfere with the analysis: carry out a

UV digestion or other suitable sample preparation.

2. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

3. If a second element is present at the same potential: add this

element to the sample and carry out the analysis again. If the
second peak has become higher then the second element is
present. Might it be possible to selectively mask this second
element with a complex former?

4. For Cu: work without chlorides in the electrolyte or increase

the chloride concentration massively.

5. Has any substance formed a precipitate in the measuring

vessel (e.g. lead perchlorate standard with KCl as electro-
lyte)?

6. Try out eluents with different compositions (addition of com-

plex formers).

7. Check method parameters.
8. Try another measurement mode like AC1. If one substance is

reversible and the second one irreversible, only the reversible
substance is detected by AC1.

With

MME:

1. Check MME. If necessary, change capillary or replace sealing

needle (see Hardware Manual).

With

RDE/SSE:

1. Check RDE and polish if necessary (see Hardware Manual).