4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

21

■

Plug in the straight plug of the working electrode connection
cable (red sleeve) into the WE socket of the detector.

■

Plug in the straight plug of the reference electrode connection
cable (black sleeve) into the RE socket of the detector.

■

Plug in the straight plug of the auxiliary electrode connection
cable (blue sleeve) into the AE socket of the detector.

3 Insert the dummy cell

■

Place the dummy cell into the tray of the detector.

■

Attach the front lid.

Note

The metal parts of the cable plugs must not touch the front lid.

4 Adjust settings in MagIC Net™

In the program part Method, create a new method for the instru-
ment test with the dummy cell.

■

Select the detector and add it as a new device.

■

Select the DC mode.

■

Set the following parameters for the DC mode:

– DC potential: 0.8 V
– Range: Auto
– Damping: off

■

Add an analysis for the detector channel Current.

■

Save the method.

In the program part Workplace:

■

Load the method.

■

In the Watch window, display the Current channel with at least
3 decimal places.

5 Carry out the test

In the program part Manual:

■

Switch on the cell
After no more than 1 minute, the detector signal should level off
at 2.667 nA ± 7 %. Noise should not exceed 0.005 nA.