6 Troubleshooting

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

37

Problem

Cause

Remedy
baseline corresponds to the one in the Appli-
cation Works.

Detector parameters –
Potentials set incorrectly.

Set the potentials to correspond to the specifi-
cations in the leaflet and in the Application
Works.

Incorrect eluent in the ref-
erence chamber.

Remove purge stopper on the measuring cell,
wait until approx. 1 mL of eluent has emerged,
screw the purge stopper back in tightly.

Electrodes contaminated.

■

Clean and polish working electrode.

■

Possibly clean the auxiliary electrode.

■

Replace the reference electrode with a
well-conditioned new reference electrode.

Very noisy baseline.

Disruptive influences from
outside.

■

In the DC mode: Switch on the damping.

■

In the other measuring modes: Set a suita-
ble smaller measuring range.

■

Attach the front lid.

Ag/AgCl reference elec-
trode worn out.

Replace the reference electrode .

Auxiliary electrode conta-
minated.

Clean the auxiliary electrode of the measuring
cell .

Working electrode conta-
minated.

■

Clean and polish the working electrode
(see the leaflet for the working electrode).

■

Replace the GC working electrode if it has
been used with oxidative potentials at the
upper limit and polishing no longer helps.

Air bubble in the measur-
ing cell.

Deaerate the measuring cell (see Chapter 4.6,
page 27)
.

Background current too
high, e.g. caused by conta-
minated eluent.

Check the background current, e.g. use fresh
eluent.