6 deaerating the measuring cell, Deaerating the measuring cell – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 35

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

27

■

Observe the measuring cell: No liquid should emerge from the
body of the measuring cell.
If the measuring cell is leaking:

– Remove the measuring cell from the measuring cell holder.
– Remove all capillaries and cables.
– Check the seating of the pressure screw of the working

electrode and retighten it.

– Restore the capillary connections.
– Reinsert the electrode cables.
– Reinsert the measuring cell.
– Repeat the test.

4 Test at normal flow

■

In the program part Manual of MagIC Net™, raise the flow of
the high pressure pump to 1.0 mL/min.

■

Observe the measuring cell: No liquid should emerge from the
body of the measuring cell.

4.6

Deaerating the measuring cell

The cell must be deaerated in order to ensure that it contains no air bub-
bles.

The measuring cell must be deaerated after the installation and after each
subsequent opening of the cell.

Proceed as follows:

Deaerating the measuring cell

Prerequisites:

■

The high pressure pump is switched on and pumps the eluent through
the IC system to the measuring cell.

■

The measuring cell is switched off.

1 Deaerate the reference electrode chamber

■

Unscrew the nut on the RE connector and remove it.

■

Lift out the reference electrode.

■

Wait until the reference electrode chamber has filled with eluent.

■

Reinsert the reference electrode. Wipe up any emerging eluent
with a cloth.

■

Screw the nut on the reference electrode connector back on
tightly.