Metrohm tiBase 1.0 Manual User Manual
Page 266

4.6 Subwindow Information
■■■■■■■■■■■■■■■■■■■■■■
258
■■■■■■■■
tiBase 1.0
Solution 'Solution name'
Solution name:
Name of the solution.
Concentration:
Concentration value and unit of the solution.
Titer:
Titer value and units of the solution.
Date titer det.:
Date of the titer determination.
Titer method:
Method with which the titer was determined.
Sensor 'Sensor name'
Information about the sensor used.
Sensor type:
Type of sensor (+ designation IS for intelligent sensors).
Sensor serial number:
Serial number of sensor.
Order number:
Order number of sensor.
Slope:
Electrode slope of the sensor used (in % for pH sensor or in mV for ISE).
pH (0):
Electrode zero point of the sensor used (dimensionless for pH sensor).
Calibration temperature:
Temperature during the calibration.
Calibration method:
Shows the method with which the calibration was carried out.
User:
Shows the user logged in during the calibration or who entered the calibra-
tion data manually.