Metrohm tiBase 1.0 Manual User Manual
Page 33

■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
tiBase 1.0
■■■■■■■■
25
Variable
name
(PC-Control)
Variable
name
( tiBase 1.0 )
Description
Command
.ESI#
.EP{x}.MEP
Marking for the endpoint x (1...9);
1 = 1 endpoint, 2 = 2 or more
endpoints
DET, MET, SET, KFT
.EP#
.EP.QTY
Measured value (water) for the
endpoint in mg
KFC
.ET#
.EP{x}.TEM
Temperature for the endpoint x
(1...9) in °C
DET, MET, SET, KFT
.ED#
.EP{x}.TIM
Time for the endpoint x (1...9) in
s
DET, MET, SET, KFT, KFC
.EP#
.EP{x}.VOL
Volume for the endpoint x (1…9)
in mL
DET, MET, SET, KFT
.MCT
.ETE
End temperature (temperature
following processing of the com-
mand) in °C
DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow), CAL MEAS, STDADD,
DOS
.MCV
.EVT
End volume (total dosed volume
at the end of the command) in mL
DET, MET, SET, KFT, STAT,
DOS
.FM#
.FP{x}.MEA
Measured value for the fixed end-
point x (1…9) in the unit of the
measured value
DET, MET, SET, KFT, STAT
.FP{x}.QTY
Measured value (water) for the
fixed endpoint x (1...9) in mg
KFC
.FT#
.FP{x}.TEM
Temperature for the fixed end-
point x (1…9) in °C
DET, MET, SET, KFT, STAT,
MEAS (without T/Flow)
.FD#
.FP{x}.TIM
Time for fixed endpoint x (1…9)
in s
DET, MET, SET, KFT, KFC,
STAT, MEAS (without T/
Flow)
.FP#
.FP{x}.[VOL]
Volume for the fixed endpoint x
(1...9) in mL
DET, MET, SET, KFT, STAT
.HM#
.HP{x}.MEA
Measured value for HNP x (1...9)
in mV
1)
DET, MET
.HT#
.HP{x}.TEM
Temperature for HNP x (1...9) in
°C
DET, MET
.HD#
.HP{x}.TIM
Time for HNP x (1...9) in s
DET, MET
.HP#
.HP{x}.[VOL]
Volume for HNP (half neutraliza-
tion potential) x (1...9) in mL
DET, MET
#M.MIM
.IME
Initial measured value (measured
value before start conditions are
processed) in the unit of the
measured value (number)
DET, MET, SET, KFT, STAT,
MEAS, CAL MEAS,
STDADD, DOS