2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

tiBase 1.0

Variable
name

(PC-Control)

Variable
name

( tiBase 1.0 )

Description

Command

.BM#

.BP{x}.MEA

Measured value for the break
point x (1…9) in the unit of the
measured value

DET, MET, MEAS (without
T/Flow)

.BT#

.BP{x}.TEM

Temperature for the break point x
(1…9) in °C

DET, MET, MEAS (without
T/Flow)

.BD#

.BP{x}.TIM

Time for the break point x (1…9)
in s

DET, MET, MEAS (without
T/Flow)

.BP#

.BP{x}.VOL

Volume for the break point x (1…
9) in mL

DE, MET

.CONC

.CONC

Concentration of the solution
used for the command (number)

DET, MET, SET, KFT, STAT,
STDADD, ADD, DOS, LQH

.MCD

.DBL

Duration command (total dura-
tion for the processing of the
command) in s

DET, MET, SET, KFT, KFC,
STAT, MEAS, CAL MEAS,
STDADD, DOS

.MDC

.DRI

Current and/or last drift for drift
correction in mL/min

SET, KFT, KFC

.MSD

.DSC

Duration start conditions (time for
processing all start conditions) in
s

DET, MET, SET, KFT, STAT

.DDC

.DTI

Time for drift correction (time
from the start of the titration to
the end of the command) in s

SET, KFT, KFC

.MCM

.EME

End measured value (measured
value after processing of the com-
mand) in the unit of the measured
value

DET, MET, SET, KFT, KFC,
STAT, MEAS, CAL MEAS,
STDADD, DOS

.MEN

.ENP

Electrode zero point for the sen-
sor used for the command
(dimensionless for pH sensor or in
mV for the ISE sensor, number)
and/or electrode zero point calcu-
lated from the calibration (for
STDADD and CAL LOOP)

DET pH, DET U, MET pH,
MET U, SET pH, SET U,
STAT, MEAS pH, MEAS U,
MEAS T, MEAS Conc, CAL
LOOP, DOS, STDADD

.EP{x}.DME

Measured value differential for
the endpoint x (1…9)

MET

.EF#

.EP{x}.ERC

ERC for the endpoint x (1…9)

DET

.EM#

.EP{x}.MEA

Measured value for the endpoint
x (1…9) in the unit of the meas-
ured value

DET, MET, SET, KFT, KFC

Advertising