7 Technical specifications
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
■■■■■■■■
99
Reproducibility
of the eluent
flow
< 0.1% deviation
Pressure range
Pump
0 - 50.0 MPa (0 - 500 bar)
Pump head
0 - 35.0 MPa (0 - 350 bar) (applies for the standard PEEK pump head)
Residual pulsa-
tion
< 1%
Safety shutdown
Function
Automatic shutdown upon reaching the pressure limit values
Maximum pres-
sure limit
■
Adjustable from 0.1 - 50 MPa (1 - 500 bar)
■
The pump is automatically shut down at the first piston stroke
above the maximum limit value
Minimum pres-
sure limit
■
Adjustable from 0 - 49 MPa (0 - 490 bar)
■
The shutdown mechanism is inactive at 0 MPa
■
The shutdown mechanism becomes active 2 minutes after system
start
■
The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit
7.8
Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
7.9
Suppressor
Resistance to sol-
vents
No restriction
Switching time
typ. 100 ms