3 removing transport locking screws, Removing transport locking screws, Figure 6 – Metrohm 930 Compact IC Flex ChS/PP/Deg User Manual
Page 25: Removing the transport locking screws

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
■■■■■■■■
17
3.3
Removing transport locking screws
To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:
■
Hex key 4 mm (6.2621.030)
Remove the transport locking screws
Figure 6
Removing the transport locking screws
1
Transport locking screws
For the vacuum pump. Only present if the
instrument has a degasser.
2
Transport locking screws
For the high-pressure pump.
1 Remove all of the transport locking screws with the hex key.
Store the transport locking screws in a safe place. Reinsert the transport
locking screws each time you transport the instrument a significant dis-
tance.