Metrohm 930 Compact IC Flex ChS/PP/Deg User Manual
Page 94

5.11 Suppressor
■■■■■■■■■■■■■■■■■■■■■■
86
■■■■■■■■
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
■
The rotor's sealing surface is approximately 4 mm inside the hous-
ing if the rotor is inserted correctly. If this is not the case, the
rotor must be moved into the right position from below using a
sharp object.
5 Cleaning the new connecting piece
■
Clean the surface of the new connecting piece (22-2) with etha-
nol using a lint-free cloth.
6 Inserting the new connecting piece
■
Insert the connecting piece (22-2) into the housing (22-4) in such
a way that the connector 1 is on top and the three pins of the
connecting piece fit into the corresponding recesses on the hous-
ing.
■
Reattach the union nut (22-1) and tighten it by hand.
7 Connecting and conditioning the suppressor
■
Reconnect all capillaries of the suppressor to the IC system.
■
Before switching the suppressor over for the first time, rinse the
three suppressor units with solution for 5 minutes.