1 installing the eluent path, 2 connecting the regeneration solution – Metrohm 930 Compact IC Flex ChS/PP/Deg User Manual
Page 42

3.11 Suppressor
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
3.11.2.1
Installing the eluent path
The eluent path is connected with the capillaries in and out.
1 Connect the capillary labeled with in with the output of the separa-
tion column using a short pressure screw (6.2744.070).
2 Connect the capillary labeled with out to the detector input capillary
using one (6.2744.040) coupling and two (6.27474.070) pressure
screws (see manual of the detector).
3.11.2.2
Connecting the regeneration solution
The regeneration solution is connected to the regenerant capillary.
Connecting the regeneration solution to the peristaltic pump
For this step you need the following accessories:
■
Accessory kit: Flex/Vario: ChS (6.5000.030)
■
Pump tubing (6.1826.420)
■
Tubing olive with filter and lock (6.2744.180)
■
Tubing olive (6.2744.034)
■
Tubing cartridge of the peristaltic pump