10 peristaltic pump, 11 detector, 12 sample degasser – Metrohm 930 Compact IC Flex ChS/PP/Deg User Manual
Page 108: 13 power connection

7.10 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
100
■■■■■■■■
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
7.10
Peristaltic pump
Type
2-channel peristaltic pump
Direction of rota-
tion
Clockwise/counterclockwise
Rotational speed
0 - 42 rpm in 7 levels of 6 rpm each
Pumping charac-
teristics
0.3 mL/min at 18 rpm; with standard pump tubing (6.1826.420)
Pump tubing
material
Recommended: PharMed® (Ismaprene)
7.11
Detector
You can find the technical specifications for the detector in the detector's
manual.
7.12
Sample degasser
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Time to establish
vacuum
< 60 s
7.13
Power connection
Required voltage
100 - 240 V ± 10% (autosensing)
Required fre-
quency
50 - 60 Hz ± 3 Hz (autosensing)
Power consump-
tion
■
65 W for typical analysis application
■
25 W standby (conductivity detector to 40 °C)
Power supply unit
■
Up to 300 W maximum, electronically monitored
■
Internal fuse 3.15 A