3.2 Operation using «IC Net 2.1»
817 Bioscan
37
3.2.2
817 Bioscan window
817 icon / Open
The
817 Bioscan
window for parameter settings is opened by
selecting this menu option with the right-hand mouse button
or by double-clicking the
817 icon
in the
SYSTEM
window. It
consists of the four tabs
Manual
,
Method parameters
,
Program
,
and
Links
.
Manual
The
Manual
tab of the
817 Bioscan
window is only available
for a connected system.
Mode
Selection of detector mode:
DC
In the
DC
mode a constant potential is ap-
plied; the analytes are oxidized or reduced
in accordance with their electrochemical
properties.
Pulse
The
Pulse
mode works with three different
potentials that are applied cyclically. This
frees the electrode surface from any adher-
ing reaction products during each sweep
and an activated surface is produced for
the next measurement.
Scan
The
Scan
mode allows potential/current
curves to be recorded.