3.2 Operation using «IC Net 2.1»
817 Bioscan
43
Scan mode Parameters
The
Scan mode
subtab of the
Method parameters
tab on the
817 Bioscan
window is used for parameter setting and curve
display for the scan mode. This window is only available if
Mode
has been set to
Scan
on the
Manual
tab.
Method data
Cell potential
Definition of start and end potential for a
user-defined potential sweep that is applied
to the amperometric detector cell.
E1
Start potential for sweep.
Range:
-2 ... +2 V
.
E2
End potential for sweep.
Range:
-2 ... +2 V
.
Scan cycle
Definition of scan cycle:
full
double
scan
E1 ... E2 ... E1
half
single
scan
E1 ... E2
Selection:
full, half
Scan rate
Definition of scan rate in mV/s.
Selection:
1, 2, 5, 10, 20, 50 mV/s
Oven
Switches column oven on/off and sets op-
erating temperature.
Range:
15 ... 60 °C
(> room temp. + 10 °C)
Range
Selection of measuring range:
pA
10 ... 5000 pA
.
nA
0.1 ... 50 nA
.
µA
0.01 ... 5 µA
.