3.2 Operation using «IC Net 2.1»
817 Bioscan
39
Method parameters
DC mode
The
DC mode
subtab of the
Method parameters
tab on the
817
Bioscan
window is used to set the parameters for the DC
mode. This window is only available if
Mode
has been set to
DC
on the
Manual
tab.
Method data
Cell potential
Electric potential applied to the amperomet-
ric detector cell in V
Range:
-2 ... +2 V
.
Oven
Switches column oven on/off and sets op-
erating temperature.
Range:
15 ... 60 °C
(> room temp. + 10 °C)
Range
Selection of measuring range:
pA
10 ... 5000 pA
.
nA
0.1 ... 50 nA
.
µA
0.01 ... 5 µA
.
Full
scale
The full-scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full-scale range
depend on the preset measuring
Range
. For
RS232 data (IC Net 2.1), always the maxi-
mum settings should be used.