6 appendix, 1 technical data – Metrohm 817 Bioscan User Manual
Page 73

6.1 Technical data
817 Bioscan
65
6 Appendix
6.1 Technical
data
Measuring unit
Operating modes
DC, Pulse and Scan
Potential
Range:
-2.00 ...+ 2.00 V
Resolution: 10
mV
Operating modes
DC mode
Range:
10 pA – 5 µA
Resolution: 1,2,5
Steps
Pulse mode
Range:
10 nA – 5 µA
Resolution: 1,2,5
Steps
Pulse times:
t1: 0.1 – 2.0 s
t2: 0.1 – 2.0 s
t3: 0 (off) – 2.0 s
Resolution: 10
ms
Sample times: 20, 40, 60, 80 and 100 ms
Scan
Range:
10 nA – 5 µA
Resolution: 1,2,5
Steps
Scan rate.:
1 – 50 mV/s
Resolution: 1,2,5
Steps
Cycle: half,
full
Autozero
Function
Shift of full scale window of RS232 data;
automatic zero setting (electronic background compensa-
tion) of the analog output
16 µA in µA range
160 nA in nA range
16 nA in pA range
Initiation
Externally (RS232, Remote)