Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual

Page 123

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Troubleshooting

850 Professional IC – Anion MSM-HC – MCS – Prep 1

■■■■■■■■

113

Problem

Cause

Remedy

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Conductivity detec-
tor is not recognized
in the software

No connection.

Check the cable connection (37-

1

).

Turn the instrument off and on again (after
15 seconds).

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Connections – dead vol-
ume in the system.

Check the connections (see Chapter 3.5, page
24)
(use PEEK capillaries with an internal diam-
eter of 0.25 mm between the injection valve
and detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 3.21,
page 72)
.

Separation column – dead
volume at column head.

Install separation column in reverse flow
direction (if permitted by the information
sheet).

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Greater rise in the
baseline

MSM-HC – reduced
capacity.

Regenerate the MSM-HC (see Chapter
5.11.3.1, page 98)
.

Precision problems -
significant scatter-
ing of the measured
values

Sample – Gas bubbles in
the sample.

Use sample degasser (see Chapter 3.13, page
50)
.

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.14.1, page 51)
.

Sample – Rinsing volume
too low.

Prolong the rinsing time (see Chapter 5.8,
page 95)
.

Injection valve – defective.

Request Metrohm Service.

MCS – vacuum too low.

Check connectors. If they are ok:

Contact the Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.15.4, page 108)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Advertising