Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual

Page 28

Advertising
background image

3.3 Installation diagram

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

3

PEEK pressure screw, short 6.2744.070

4

MSM-HC eluent inlet capillary
Labeled with Eluent. Connected to the col-
umn output.

5

MSM-HC eluent outlet capillary
Labeled with Detector. Connected to the
input of the MCS.

6

PEEK pressure screw, long 6.2744.090

7

Conductivity detector

8

Detector inlet capillary
Connected to the output of the MCS.

9

Detector outlet capillary

10 MSM-HC regeneration solution aspirat-

ing capillary 6.1803.020

11 Tubing olive 6.2744.034

For connecting capillaries to the aspiration
side of the peristaltic pump.

12 Pump tubing 6.1826.320

With orange/yellow stoppers. For the regen-
eration solution.

13 Pump tubing connection 6.2744.180

With safety device and filter. For connecting
capillaries to the outlet side of the peristaltic
pump.

14 MSM-HC regeneration solution inlet

capillary
Labeled with H2SO4.

15 MSM-HC regeneration solution outlet

capillary
Labeled with Waste.

16 MSM-HC rinsing solution aspirating

capillary 6.1803.020

17 Pump tubing 6.1826.320

With orange/yellow stoppers. For the rinsing
solution.

18 MSM-HC rinsing solution inlet capillary

Labeled with H2O.

19 MSM-HC rinsing solution outlet capil-

lary
Labeled with Waste.

20 MCS adsorption cartridges

21 Air aspirating capillary

For aspirating CO2-low air from the car-
tridges of the MCS.

22 Separation column

Dialysis
The following installation diagram shows the connection of the sample
path with dialysis as inline sample preparation. For this you need the
6.5330.000 IC equipment for dialysis to be ordered separately.

For dialysis, an acceptor solution is preconcentrated with the ions of the
matrix-loaded sample and is carried via the injection valve onto the sepa-
ration column. The peristaltic pump in the instrument pumps the acceptor
solution; the sample still heavily matrix-loaded is pumped with a second
peristaltic pump which is installed on the Sample Processor. The tubing of
the peristaltic pump on the Sample Processor is not shown in this dia-
gram.

Advertising