3 Installation

850 Professional IC – Anion MSM-HC – MCS – Prep 1

■■■■■■■■

15

Connect the capillary labeled with H20 with a 6.2744.180 pump
tubing connection and a 6.2744.070 short pressure screw to the
peristaltic pump on the outlet end of the pump tubing for the
rinsing solution.

Guide the two capillaries labeled with Waste into a waste con-
tainer and fasten it there.

7 Connecting the instrument

Connect the instrument to the PC with a 6.2151.020 USB cable
(see Chapter 3.20.1, page 71).

Connect instrument to mains supply (see Chapter 3.20, page
71)
.

8 Initial start-up

(see Chapter 4.1, page 76)

Switch on the PC and start MagIC Net.

Switch on the instrument.

Deaerate the high pressure pump.

Set the contact pressure of the peristaltic pump

Rinse instrument without column.

9 Installing guard and separation column

Remove the 6.2744.040 coupling between the column inlet capil-
lary and the eluent inlet capillary of the MSM-HC.

Connect guard column (optional) (see Chapter 3.21, page 72)

– Fasten the guard column to the end of the column inlet

capillary according to the specifications contained in the
leaflet accompanying the guard column.

– Rinse guard column.

Connecting the separation column (see Chapter 3.22, page 74)

– Fasten the inlet of the separation column either to the end

of the column inlet capillary or to the guard column (if
used) according to the specifications contained in the leaf-
let accompanying the column.

– Fasten the MSM-HC capillary labeled with Eluent with a

6.2744.070 PEEK pressure screw at the output of the sepa-
ration column.

Hang separation column with chip in the column holder of the
instrument.

10 Conditioning the instrument

(see "Conditioning the system", page 77)