Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual

Page 24

Advertising
background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

3 Connecting the eluent path

Equip the 6.1834.080 eluent aspiration tubing and connect with
the eluent bottle (see Chapter 3.8.1, page 37).

Connect the 6.1831.150 column inlet capillary and the capillary of
the MSM-HC labeled with Eluent with a 6.2744.040 coupling and
two 6.2744.070 short pressure screws.

Use a 6.2744.090 long pressure screw to connect the capillary of
the MSM-HC labeled with Detector to the input of the MCS (see
Chapter 3.18.2, page 65)
.

Connect the detector inlet capillary with a 6.2744.090 long pres-
sure screw to the output of the MCS (see Chapter 3.18.2, page
65)
.

4 Connecting the sample path

Connect the sample aspirating capillary and the sample outlet capil-
lary according to the instructions in the manual of the IC equipment
used:

6.5330.000 IC equipment for dialysis
(see manual IC equipment – 6.5330.000 – Dialysis).

6.5330.010 IC equipment for ultrafiltration
(see manual IC equipment – 6.5330.010 – Ultrafiltration).

5 Installing the peristaltic pump

(see Chapter 3.17.2, page 61).

Connect the 6.1803.020 aspirating capillary for regeneration solu-
tions with a 6.2744.034 tubing olive and a 6.2744.070 short
pressure screw to the aspiration end of the 6.1826.320 pump
tubing for the regeneration solution.

Place the pump tubing in a tubing cartridge.

Connect the 6.1803.020 aspirating capillary for rinsing solutions
with a 6.2744.034 tubing olive and a 6.2744.070 short pressure
screw to the aspiration end of the second 6.1826.320 pump tub-
ing for the rinsing solution.

Place the second pump tubing in the second tubing cartridge.

Place both tubing cartridges into the peristaltic pump.

6 Connecting the MSM-HC

(see Chapter 3.16.2, page 57)

Connect the capillary labeled with H2SO4 with a 6.2744.180
pump tubing connection and a 6.2744.070 short pressure screw
to the peristaltic pump on the outlet end of the pump tubing for
the regeneration solution.

Advertising