3.16 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
850 Professional IC – Anion MSM-HC – MCS – Prep 1
1
3
2
3
4
5
6
7
8
2
1
Figure 29
MSM-HC – Connectors
1
Union nut
2
MSM-HC connecting piece 6.2835.010
3
Eluent inlet capillary
Labeled with Eluent.
4
Eluent outlet capillary
Labeled with Detector.
5
Rinsing solution inlet capillary
Labeled with H2O.
6
Rinsing solution outlet capillary
Labeled with Waste.
7
Regeneration solution outlet capillary
Labeled with Waste.
8
Regeneration solution inlet capillary
Labeled with H2SO4.
The PTFE capillaries firmly mounted on the MSM-HC are connected to the
other components of the IC system as follows:
Caution
As the PTFE capillaries are very soft, the pressure screws should not be
overtightened.
Flattened capillaries can be shortened with the help of a capillary cutter.