6: Action line

A step in the method sequence, reflecting the values entered by selecting
or typing values in the Pump, Valve, and Wait fields. To move to a differ-
ent line for editing, just click on it or use

No. 12: Up and Dn buttons.

7: Pump function

Aspirate:

Draw liquid into the line connected to the pump outlet (Port B).

Dispense:

Force a pre-programmed amount of liquid through the pump.

Wash:

Wash or flush the lines (always in the outward direction,

or A to B) with system fluid or a pre-programmed reagent

selected.

8: Volume

The amount to be aspirated, dispensed, or sent through as a wash.

9: Flow rate

The rate at which the given volume is aspirated, dispensed, or sent
through as a wash.

10: Wait field

If a value is entered in this field for an action line, during a run the
program will wait for the designated time (in seconds) after the
completion of that action before advancing to the next action. Used
after an aspirate or dispense step to allow the liquid to equilibrate in the
line or the dispensed liquid to come to equilibrium, as with viscous
liquids or very small capillary tubing.

11: Event box

When this box is clicked when programming an action line, the program
will stop before that action is executed. The program will remain stopped
until

No. 4: Event Button is clicked. (If the step before the event line

has a programmed Wait interval, the Event button will not trigger the
action until the Wait interval has elapsed.)

12: Up and D(ow)n buttons

In the Editing mode, these move the line highlighter up and down.

13: Valve type

Multi-Pos: Multiposition stream selector
2-Pos:

Two position switching valve/injector

14: Valve Position

The position the valve needs to be in to select the proper solvent or
reagent.

15: Valve Direction (for Multi-Pos only)

The direction the valve will take to the indicated position – the shortest
(used in most cases), clockwise, or counterclockwise.

16: Insert button

Inserts a new action line above the highlighted line.

17: Duplicate button

Duplicates the highlighted line. The new line can then be edited as required.

18: Delete button

Deletes the highlighted line.

19: End Edit button

When editing is complete, click this button to return to the Idle mode,
ready to Run.