VICI M Series User Manual

Page 27

Advertising
background image

23

Command

Description

MHC=n

Motor holding current used when the pump is stationary, n=% of 4A

MRC=n

Motor run current used when the pump is running, n=% of 4A

HCDT=n

Delay time in milliseconds between cessation of motion and the return to

the current level specified by the MHC variable

MSEL=n

Micro stepping resolution, n=2,4,8,16,32,64,128,256

MUNIT=n

Conversion factor from motor pulses to user units (µL), n=19 for MSEL=2,

n=304 for MSEL=32. MUNIT=

BLE=n

Backlash compensation enable flag, n=1 enable, n=0 disable

BLM=n

Backlash mode, n=0 mechanical, n=1 mathematical

BLSH=n

Backlash compensation amount, µL

CTR1

Counter of pulses sent to motor

ALL

Variable used with GET, PRINT, and IP instruction

ERR

Read only flag that indicates whether an error has occurred

ERROR

Read only variable that gives the error code of the most recent error

FLAGS

Used with GET, PRINT, and IP instruction

IP n

Initialize specified parameters to factory default settings, n omitted initializes

ALL, n=ALL, VARS, FLAGS, IOS. You will need to manually update

variables marked with an *.

PGM

Keyword to place MicroLynx into program mode

PRINT n

Instruction to output variable value to screen

SAVE

Instruction to save ALL variables to non-volatile memory

SER

Read only variable that contains the serial number of the MicroLynx

VARS

Used with GET, PRINT, and IP instruction

VER

Read only variable that contains the present MicroLynx firmware version

CP=n

Clears the program space on the MicroLynx

EXEC n

Execute the program located at address n

LIST n

List the program starting at address n

<ESC>

Terminate all active operations and running programs

Ctrl-C

Terminate all active operations and running programs, forces a partial reset

BAUD=n

Sets the baud rate

DN="n"

Set the device address for PARTY mode operation

ECHO=n

Specifies whether commands will be echoed back over comm line, n=0
Full duplex, n=1 Half duplex, n=2 Only respond to PRINT and LIST commands

PRMT

Specifies the character to be used as a prompt character

BSY

Flag that indicates whether a program is running

PARTY

Enables / disables PARTY mode for multi drop applications

HOST

Establish or disble a controller as host in a multi-drop installation.

Host=1 controller is host, Host=0 controller is not host. Host controllers

receive commands on RS-232 and transmit them to other units via RS-485

Factory

96

“!”

0

>

0*

35*

256*

2432 *

1*

1*

1.5**

0*

**

*

Indicates variables which must be manually updated after an IP command.

*

The backlash value must be manually set to the value on the QC certificate.

MSEL step resolution * (360/motor step size) * gear ratio

volume in µL of one revolution

Figure 15: MicroLYNX-4 commands, part B

We recommend that you contact VICI technical support
before changing any of the factory default values. They
are application dependent, and any changes could ad-
versely affect the pump and its performance.