Appendix b: setup for multipump operation, Connections, Microlynx #1 microlynx #2 microlynx #n – VICI M Series User Manual

Page 28

Advertising
background image

24

7 pin connector

Signal

4

RX+

3

RX-

5

TX-

7

TX+

6

C Gnd

7 pin connector

Signal

7

TX-

5

TX+

3

RX+

4

TX-

6

C Gnd

7 pin connector

Signal

7

TX-

5

TX+

3

RX+

4

TX-

6

C Gnd

Host=1
Party=1
Echo 2=0

Set system variables to the following values:

Set the device address using the address DIP switches:

Address: A
A0=ON, A1=OFF, A2=OFF

Host=0
Party=1
Echo 2=0

Address: B
A0=OFF, A1=ON, A2=OFF

Host=0
Party=1
Echo 2=0

Address: C
A0=OFF, A1=OFF, A2=ON

MicroLYNX #1

MicroLYNX #2

MicroLYNX #n

Four wire RS-485

7 pin connector

Signal

4

B

3

A

5

A

7

B

6

C Gnd

7 pin connector

Signal

7

A

5

B

3

B

4

A

6

C Gnd

7 pin connector

Signal

7

A

5

B

3

B

4

A

6

C Gnd

Host=1
Party=1
Echo 2=1

Set system variables to the following values:

Set the device address using the address DIP switches:

Address: A
A0=ON, A1=OFF, A2=OFF

Host=0
Party=1
Echo 2=1

Address: B
A0=OFF, A1=ON, A2=OFF

Host=0
Party=1
Echo 2=1

Address: C
A0=OFF, A1=OFF, A2=ON

MicroLYNX #1

MicroLYNX #2

MicroLYNX #n

Two wire RS-485 (e.g., to couple to valves as well)

Appendix B: Setup for Multipump Operation

Connections

Make up cables using the following guidelines. Note that communication
with the PC is assumed to be via RS-232.

If you are using an RS-232 to RS-485 converter, treat the converter as the
Host, with all MicroLYNXes configured as slaves.