Operational and technical specifications, M6 m50 – VICI M Series User Manual

Page 25

Advertising
background image

21

Physical Specifications
Dimensions

See Figure 2

See Figure 2

Weight: pump

520 g

540 g

motor and gear assy.

560 g

650 g

micro-electric controller

300 g

300 g

power supply

300 g

300 g

Resolution

MSEL=256

MSEL=256

Minimum flow

5 nL/min

1 µL/min

Maximum continuous flow

5 mL/min

25 mL/min

Maximum intermittent flow*

10 mL/min

35 mL/min

Maximum back pressure

100 psi

100 psi

Volume accuracy (%error)

±0.5% at 1.25 mL

±0.5% at 8 mL

±0.5% at 125 µL

±0.5% at 800 µL

Volume precision (%CV)

<0.1% at 1.25 mL

<0.1% at 8 mL

<0.5% at 125 µL

<0.5% at 800 µL

Fittings

1/16" Valco detail (10-32) 1/8" Cheminert (1/4-28)

Fluid contact

PTFE, Sapphire, PAEK,

PTFE, Ceramic,

Viton

®

, and Valcon P

Valcon H2

Planetary Gear Assembly
Ratio

1:4.75

1:9.88

Motor
Type

2 Phase Bipolar (1.8° per step)

Environmental
Operating Temperature

0 - 50°C

Operating Humidity

< 90% non-condensing

Storage Temperature

-20°C – 70°C (dry)

Communications (COM1 or COM2)
Type

RS-232 standard, RS-485 (RS-422) optional

Baud rate

Default 9600 bps, Full range 4800 - 38,800 bps

Parity

None

Data bits per character

8

Stop bits

1

Error checking

None

Mode

ASCII

Power requirement
Rest current draw

0 mA

Operational current draw

Typical 800 mA, 1600 mA at acceleration

Voltage

Typical 24 VDC, input voltage 12 - 48 VDC

Both Models (M6 and M50)

M6

M50

* High speed continuous flow or the use of chemicals incompatible with the wetted surfaces

*

(refer to page 17) will reduce pump lifetime.

Operational and Technical Specifications