Bio-Rad PDS-1000 / He™ and Hepta™ Systems User Manual

Page 33

Advertising
background image

Fig. 4.13. View of spent rupture disk within the retaining cap.

4.3 Removal of Residual Helium Pressure—Shut Down

After completing all bombardment(s), remove the helium pressure from the PDS-1000/He

system and close the helium cylinder valve. Perform the following steps to remove helium

pressure from the system.
1. Close the helium cylinder valve and chamber door.
2. With at least 5 inches of Hg of vacuum in bombardment chamber, remove residual line

pressure from the regulator, solenoid and PEEK tubing by activating the FIRE button.

3. Release the FIRE button on the apparatus, and remove all tension on the pressure adjust-

ment screw of the helium regulator, turning counter-clockwise, until it turns freely.

4. Vent any residual vacuum from the bombardment chamber by setting the vacuum switch

to the VENT position.

Section 5
Selection and Adjustment of
System Bombardment Parameters

5.1 Overview—Matrix of Variables

Factors affecting bombardment efficiency are numerous, and interact in complex ways.

All possible variables cannot be addressed in a single large factorial experiment. Instead, most

users will find it sufficient to prioritize variables by the magnitude of their effects, optimizing

them individually, and then testing their interactions on a more limited scale. Table 5.1 shows

some examples of conditions used in published protocols for a variety of target materials.

30

Advertising