Bio-Rad PDS-1000 / He™ and Hepta™ Systems User Manual

Page 42

Advertising
background image

6.6 Unit Will Not Pressurize Gas Acceleration Tube

Probable Cause

Action

1. Rupture disk not in position or

Insert disk properly

not sealed by retaining cap.

2. Main cylinder valve of Helium

Adjust according to Section 3.1

tank not open

3. Insufficient vacuum

Examine vacuum connections (see Section 3.2)

4. Helium system leak

Isolate leaks in the helium pressure regulator by

closing the helium cylinder valve and turning the

pressure-adjusting screw one turn counter-clock-

wise, then note the following:
• If the high pressure gauge reading drops, there is

a leak in the cylinder valve, inlet fitting, or high

pressure gauge.

• If the low pressure gauge drops, there is a leak in

the valve, hose, hose fitting, outlet fitting or low

pressure gauge.

• If the high pressure gauge drops, and at the same

time the low pressure gauge rises, there is a leak in

the regulator seat.

Defective regulators should be returned to Bio-Rad

Laboratories for repair. Use of a damaged regula-

tor, or one that has been repaired improperly is a

safety hazard.

5. Helium tank regulator not adjusted Adjust according to Section 3.1

correctly

6. Solenoid 3-way metering valve

Connect according to Section 3.1 and 5.4

assembly not properly connected

to main unit

7. Solenoid 3-way metering valve closed Adjust according to Section 3.1 and 5.4

39