Bio-Rad PDS-1000 / He™ and Hepta™ Systems User Manual

Page 46

Advertising
background image

Section 8
Appendices

8.1 Cleaning the PDS-1000/He Device

1. Chamber

Clean the chamber with 70% ethanol. Allow time for drying.

2. Gas Acceleration tube

A. Unscrew the rupture disk retaining cap from the gas acceleration tube
Do not autoclave or flame sterilize the PDS-1000/He. Autoclaving will wet electrical

components, thus creating a potential electrical shock hazard.
B. Disconnect the tubing assembly from the gas inlet on the rear of the acceleration

tube (See Figure 1.2). Remove the screws from the hold-down brackets on the top

of the PDS-1000/He. Two sets are located on either side of the acceleration tube,

and CAREFULLY remove the tube from the top of the sample chamber.

3. The acceleration tube can be sterilized by partially immersing in 70% ethanol followed

by drying in a sterile environment. Do not autoclave the gas acceleration tube. Do not

immerse the pressure gauge in the sterilizing solution.

8.2 Metal Case Version

The metal chamber design has been offered by Bio-Rad Laboratories since the product

line was acquired from E.I du Pont de Nemours in 1991. The original metal case design is

capable of operating either the gunpowder acceleration mechanism or the helium acceleration

mechanism (Figure 8.1). The gunpowder acceleration delivery system has been discontinued.

The current plastic case version of the PDS-1000/He unit was introduced in March 1995 and

only operates with helium. The target distances and internal dimensions of the plastic case are

equivalent to those found in the metal case when used with helium acceleration mechanism.

Fig. 8.1. Front view of Metal Case version of PDS-1000/He instrument.

43