Bio-Rad Criterion™ XT Tris-Acetate Precast Gels User Manual

Page 18

Advertising
background image

Section 7
Troubleshooting

Improper storage of Criterion XT gels can produce numerous artifacts. Criterion XT Bis-Tris gels should be
stored flat at ambient temperature. Criterion XT Tris-acetate gels should be stored flat at 4°C. Avoid freezing. If
you suspect your gels have been stored improperly, DO NOT USE THEM.

Problem

Possible Cause

Solution

Samples do not migrate into gel

Tape at the bottom of the cassette

Remove tape

not removed

Insufficient buffer in integral buffer chamber

Fill integral buffer chamber with 60 ml
running buffer

Insufficient lower electrode buffer

Fill both halves of the lower buffer tank with
400 ml running buffer when running two gels

Electrical disconnection

Check electrodes and connections

Bands “smile” across gel, band pattern

Excess heating of gel

Check buffer composition

curves upward at both sides of the gel

Completely fill both halves of the lower buffer
tank with 400 ml running buffer when running
two gels

Do not exceed recommended running
conditions

Skewed or distorted bands, lateral

Excess salt in samples

Remove salts from sample by dialysis or

band spreading

desalting column prior to sample
preparation

Insufficient sample buffer

Check buffer composition and dilution

or wrong formulation

instructions

15

4110130B.qxd 3/25/2003 2:19 PM Page 15