Problem

Possible Cause

Solution

Vertical streaking

Samples overloaded

Dilute sample

Selectively remove predominant protein in
the sample

Sample precipitation

Centrifuge samples to remove
particulates prior to sample loading

Gels run too fast, provide poor resolution,

Running buffer is too concentrated

Check buffer composition

and gel temperature is too high

Artifact bands at ~60–70 kD

Possible skin keratin contamination

Wear gloves while cleaning all dishware
and while handling and loading gel

Filter all solutions through nitrocellulose

Use 10% iodoacetamide to eliminate
keratin bands