Bio-Rad PowerPac™ Universal Power Supply User Manual

Page 12

Advertising
background image

Step Procedure

Description

4.

Set the Power Failure

Use the PFd soft key to set the Power Failure

detection mode

detection mode:

PFd = OFF: Turns power failure detection
off. In this mode, the run is terminated if a
power failure occurs.
PFd = ON NEXT RUN ONLY: Turns
power failure detection on for a single run.
PFd = ON: Turns power failure detection
mode on. If there is a power failure in this
mode, then the run will resume when the
power is restored after single or mulitple
power failures.

Warning: Turning the power supply off to stop a
run in progress is regarded as a power failure and
the interrupted run (when PFd = ON) will resume
the next time the power supply is turned ON.

5.

Set the Rapid

Press the RRCd-NLDd soft key to change the

Resistance Change

Rapid Change detection and No Load

detection/No Load

detection mode.

detection mode

RRCd-NLDd = ON: Detection of a load
change greater than 20% or 15 mA, or the
absence of a load (current is less than 2
mA), will cause an alarm to sound and the
run to pause.
RRCd-NLDd = OFF: Does not detect
sudden load changes or the absence of a
load. When running in this mode, the
symbol is displayed on the Run
screen as a reminder that this safety
feature has been turned off.

Note: This mode is used to complete
electrophoresis application such as the D-Code
that require that Rapid Resistance Change
detection be turned off or applications such as
isoelectric focusing that require less than 2 mA
current.

6.

Set the control panel

Press the PANEL soft key.

default settings

7.

Set the default contrast

Press the CONTRAST soft key and enter a
value from the keypad. The contrast value range
is between 80 and 110.

!

!

6

PFd: OFF

RRCd-NLDd: ON

SETUP

PFd

RRCd-NLDd

PANEL

PFd: OFF

RRCd-NLDd: ON

SETUP

PFd

RRCd-NLDd

PANEL

Tue

CONTRAST: 95

KEY CHIRP: ON

27-May-03

CLOCK CONTRAST CHIRP

11:22