Step

Procedure

Description

11.

View the run

The "Run Completed" or "Run Terminated"

information

screen is displayed at the end of a run. The
date and time, the final running parameter
values, the total elapsed time in hours and
volt-hours are displayed on the screen. Press
the RERUN soft key to repeat the run or the
EXIT soft key to display the Method Step List.
After the run is finished, the run data may be
transmitted to a PDA or PC (see Section 4.2.2).

3.3.3 Edit a Paused Run

This section describes how to edit a paused run.

Step

Procedure

Description

1.

Pause the run

Press the PAUSE soft key or

to pause a

run that is in progress.

2.

Edit the method

The pause window displays the current run
parameters as well as the date and time that
the run was paused.

Press the EDIT soft key or the EDIT key to
modify the method. The CONTINUE soft key
or

button can be used to resume the run.

3.

Select a method step

Press the up and down arrow keys to select
the step to edit or add. An asterisk (*) indicates
the selected step.

Press the EDIT key to modify the selected
step.

Press the CONTINUE soft key or

to

resume the run when finished
editing the method.

4.

Set the constant

Use the soft keys to select the constant

parameter for the

parameter and the keypad to enter its value.

current step

Press the EDIT key when finished. The
key can be pressed at any time to
resume the run. Note that the non-constant
parameter
values are listed on the right side of the screen
and the time mode at the bottom.
ConstV:

Used to run with constant voltage.

ConstA:

Used to run with constant current.

ConstW: Used to run with constant power.

13

Time: 00:00 Vh: 00000

RUN PAUSED AT:

05/27/03 12:29:00

100V 53mA 5W

CONTINUE

EDIT

METHOD: SDS-PAGE

S1 100V 00:30 HRS

S2 300V 00:30 HRS

S3 (EMPTY)

*

CONTINUE

TIME: 00:30 HRS

S1/2 EDIT CONSTANT

CONSTA CONSTW

2500

500

mA

W

CONST

V

100 V

RUN COMPLETED AT:

05/27/03 12:34:01

200V 106mA 21W

Time: 00:01 Vh: 00003

RERUN

EXIT

00:00/00:30 00:00 HRS

PAUSE

53

5

mA

W

100 V

RUN

: S1/2