3.3.2 Edit (or View) and Run a Saved Method

This section describes how to edit, view and run a saved method.

Step

Procedure

Description

1.

Set up the power supply

Refer to sections 3.1 for instrument setup.

2.

Select the programming

On the Home screen, press the METHODS

mode

soft key to view the method list.

3.

Open an existing method Use the up and down arrow keys to select a

method to edit (identified by *) and then press
OPEN to open the Method screen. Press
to run the opened method or go to Step 4
to view and edit the method.

4.

Select a step to edit

Press the up and down arrow keys to select a
step. An asterisk (*) indicates the selected
step. Press EDIT to view or edit. Note: to edit
the method name, use the up arrow key to
move to the method name and activate the
name field. Use the keypad to change the
name.

5.

Set the constant

Use the soft keys to select the constant

parameter for the

parameter and the keypad to enter its value. If

current step

the method is complete skip to Step 9,
otherwise, press the EDIT key. Note that the
non-constant parameter values are listed on
the right side of the screen and the time mode
at the bottom.
ConstV:

Used to run with constant voltage.

ConstA:

Used to run with constant current.

ConstW: Used to run with constant power.

11

METHOD LIST

IEF

NATIVE

SDS-PAGE

*

OPEN

DELETE

NEW

METHOD: SDS-PAGE

S1 100V 00:30 HRS

S2 300V 00:30 HRS

S3 (EMPTY)

*

METHODS

DELETE

TIME: 00:30 HRS

S1/2 EDIT CONSTANT

CONSTA CONSTW

2500

500

mA

W

CONST

V

100 V

SELECT MODE

MANUAL

METHODS