Step Procedure

Description

4.

Edit a step

a.

Set the Constant mode: voltage (V),
current (A) or power (W).

b.

Enter the constant and non-constant
values in the Volts, mAmps and Watts
fields.

c.

Select whether the run will be timed
(Untimed box is not checked) or
un-timed (the Untimed box is checked).

d.

Change the method name as needed.
Check the Rename box to rename the
method or leave it un-checked to save
the edited method with the new name.

e.

Select another step to edit from the
"Step" drop-down list and edit the step.

f.

Press Save to save the method.

Other Buttons
Delete: Deletes the entire method.
Delete Step: Deletes the current step.
Methods: Returns to Method List screen.

5.

Transmit the method

Press Beam to transmit the method to the
PowerPac Universal. Note that if nine
methods already exist on the PowerPac
Universal that the new method will be
transmitted to flash memory so it can be run,
but it will not be saved to the method list.

4.2.2 Receive and Transmit Data

Step Procedure

Description

1.

Start the software

Start the PowerPac Remote software and
press Run Data

17

Method: SDS-PAGE

Step # 1

Const Mode

Rename

Volts: 100

mAmps: 2500

Watts: 500

Vhours: 00000

Untimed

Save

Delete

Beam

Methods

Delete Step

V

A

Timing

T

Vh

W

Bio-Rad Laboratories

PowerPac Remote Control

Select/Edit Methods...

Run Data