Bio-Rad PowerPac™ Universal Power Supply User Manual

Page 27

Advertising
background image

5.2.4 Error Messages

Error Message

Cause

Solution

ERROR STOP CODE 01

1. Electrophoresis cell

1. Make sure all electrical

NOLOAD DETECTED

not connected to the

connections make good

CORRECT AND CONTINUE

power supply or

contact and the cables

RUN PAUSED

buffer levels too low.

and wire electrodes are

CONTINUE

CANCEL

in good shape.

Verify buffer levels are
appropriate.

2. The current load is

2. Verify the electrophoresis

less than 2 mA.

application power
requirements match
PowerPac Universal output
range. Note: The No Load
detection can be de-
activated. (See Sections 3.1
and 5.2.3)

ERROR STOP: CODE 02

1. Accidental shorting

1. Make sure all electrical

OVER CURRENT

of output leads.

connections make good

CORRECT AND CONTINUE

contact and the cables

RUN PAUSED

and wire electrodes are in

CONTINUE

CANCEL

good shape.

2. Shorting due to

2. Verify buffer levels are

wrong connections.

appropriate.

ERROR STOP: CODE 03

1. Control circuitry

1. Press the CONTINUE soft

OVER VOLTAGE

problem

key to continue the run.

CORRECT AND CONTINUE

If the problem continues,

RUN PAUSED

cycle power and restart the

CONTINUE

CANCEL

run. If the problem persists,
contact Bio-Rad Technical
Support.

ERROR STOP: CODE 07

AC Power was interrupted To continue a run after a

RUN POWER FAILURE

during the run and the

power failure, activate the

RUN TERMINATED

run was terminated

Power Failure Detection mode

RESET WHEN READY

because the Power

in the Setup window (See

RESET

failure detection is OFF

Section 3.1).

ERROR STOP: CODE 08

Control circuitry

1. Press the RESET soft key to

REGULATION ERROR

problems.

clear the screen, then

RUN TERMINATED

restart the run.

RESET WHEN READY

2. If the problem continues,

RESET

cycle power and restart the
run. If the problem persists,
contact Bio-Rad Technical
support.

21