Step Procedure

Description

7.

Enter non-constant

Use the soft keys to select a non-constant

parameter limits for

parameter and the keypad to enter its value.

the current step

If the method is complete and will be run
un-timed go to Step 10, otherwise, press the
EDIT key. Note: to change the constant value,
press its soft key and enter a new value.

ConstV:

Used to change the voltage
constant value.

LimitA:

Used to change the maximum
default value for current limit.

LimitW:

Used to change the maximum
default value for power limit.

8.

Select the time mode

Use the soft keys to select a time mode and
the keypad to enter the run time. Press EDIT.

HRS:

Run time is in units of hours.

VHOURS: Run time is in units of volt-hours.
UNTIMED: Run is continuous. Note that steps

programmed after an un-timed
step are ignored.

9.

Add a new step or save

To program additional steps press the down

the method

arrow key to select the empty step and repeat
steps 5–8.

If the method will not be saved, go directly to
Step 10, otherwise press the SAVE soft key
and then go to Step 10. Press the DELETE
soft key to delete the selected step. Press the
METHODS soft key to return to the Method
List screen.

10.

Start the run

Press

to start a run.

11.

Monitor the run

The run screen is used to monitor run
parameters or to pause a run. The current
step versus total steps programmed, voltage,
current and power values, elapsed time versus
the programmed time, and total elapsed time
of the method are displayed on the run screen.

Press the PAUSE soft key to edit the run
parameters (see Section 3.3.3). Press

to

terminate the run. Note that any edits
made during the run will not be saved unless
SAVE is pressed at the end of the run.

12.

View the run

The "Run Completed" or "Run Terminated"

information

screen is displayed at the end of a run. The
date and time, the final running parameter
values, the total elapsed time in hours and
volt-hours are displayed on the screen. Press
the RERUN soft key to repeat the run or the
EXIT soft key to display the Method Step List.
After the run is finished, the run data may be
transmitted to a PDA or PC (see Section 4.2.2).

10

METHOD: NAME

S1 100V 00:30 HRS

S2 (EMPTY)

*

METHODS

DELETE

SAVE

00:00/00:30 00:00 HRS

PAUSE

53

5

mA

W

100 V

RUN

: S1/2

RUN COMPLETED AT:

05/27/03 12:34:01

200V 106mA 21W

Time: 00:01 Vh: 00003

RERUN

EXIT

TIME: UNTIMED

S1/1 EDIT LIMITS

LIMITW

2500

500

mA

W

CONSTV

100 V

LIMIT

A

TIME: UNTIMED

S1/1 EDIT TIME

UNTIMED

2500

500

mA

W

HRS

100 V

VHOURS