Step

Procedure

Description

5.

Set the non-constant

Use the soft keys to select a non-constant

parameter limits for

parameter and the keypad to enter its value.

the current step

Press the EDIT key when finished. The
key can be pressed at any time to
resume the run. Note: to change the constant
value, press its soft key and enter a new
value.
ConstV:

Used to change the voltage
constant value.

LimitA:

Used to change the current limit
value.

LimitW:

Used to change the power limit
value.

6.

Select the time mode

Use the soft keys to select a time mode and
the keypad to enter the run time. Press the
EDIT key when finished and go to Step 3. The

key can be pressed at any time to resume

the run. Note that the time mode

cannot be changed for the step in progress.
HRS:

Run time is in units of hours.

VHOURS: Run time is in units of volt-hours.

UNTIMED: Run is continuous. Note that steps

programmed after an un-timed
step are ignored.

7.

Exit the run completed

After the run is finished the "Run Terminated"

screen

or "Run Completed" screen is displayed. At
this point the run data may be transmitted to a
PDA or PC, if desired (see Section 4.2.2).

Press the RERUN soft key to repeat the run or
the EXIT soft key to return to the Method Step
List screen.

8.

Save the changes

From the Method Step List screen, press the
SAVE soft key to save the changes made
during the run. Note that pressing SAVE will
overwrite the original method.

14

TIME: 00:30 HRS

S1/2 EDIT TIME

2500

500

mA

W

100 V

RUN COMPLETED AT:

05/27/03 12:34:01

200V 106mA 21W

Time: 00:01 Vh: 00003

RERUN

EXIT

METHOD: SDS-PAGE

S1 100V 00:30 HRS

S2 300V 00:30 HRS

S3 (EMPTY)

*

METHODS

DELETE

SAVE

TIME: 00:30 HRS

S1/2 EDIT LIMITS

LIMITW

2500

500

mA

W

CONSTV

100 V

LIMIT

A