GS-900 Densitometer Technical Specifications

User Guide | 15

GS-900 Densitometer Technical Specifications

Scannable Sample Types

Transparent

Polyacrylamide gels

Film

Reflective

Blotting membranes

Software Specifications

Supported file types

Import: .scn, .1sc

Export: .tif, .jpg, .bmp, .png

Hardware Specifications

Image sensor type

Tri-linear CCD with three filters (red, green, blue)

Resolution

Spot size (

μm): 36.3, 42.3, 63.5, 95.3, 127.0

Light source

LED

Scanning area

Transparent: 29.0 x 33.0 cm

Reflective: 29.0 x 33.0 cm

Dynamic range

Transmissive: 0.0–3.4 OD (specular)

Reflective: 0.0–2.7 OD (specular)

Bit depth

Input: 16-bit gray scale

Output: 16-bit gray scale

Calibration

Transmissive: To internal NIST-traceable transmissive
reference standard

Reflective: To internal reflective reference standard

Connectivity

USB 2.0

Operating voltage

100–240 V AC, 47–63 Hz

Power consumption

54.0 watts

Operating temperature

10–28°C

Operating humidity

<70% noncondensing

Dimensions

Length: 63 cm

Height: 19 cm

Width: 38 cm

Weight

16 kg (35.4 lbs)