Calibrating the gs-900 densitometer – Bio-Rad Image Lab™ Software User Manual
Page 28

2 | Setting Up the Instrument with Image Lab Software
28 | GS-900 Calibrated Densitometer with Image Lab Software
Calibrating the GS-900 Densitometer
Before you can use Image Lab with the GS-900 densitometer, Image Lab must
calibrate the standards location and acquire the flat images. When you first connect
the instrument to the computer on which you are running Image Lab, the instrument
automatically begins this calibration process.
Note: Ensure that the top of the GS-900 densitometer is lowered before you
start Image Lab for the first time. If it is up, the initial calibration process will fail.
To calibrate the GS-900 densitometer
1.
Start Image Lab.
The Instrument Calibration dialog box appears, identifying the instrument to
which your computer is connected and prompting you to proceed with the
calibration.
2.
Click Next.