Calibrating the GS-900 Densitometer
User Guide | 29
The first Instrument Calibration dialog box appears prompting you to calibrate
the GS-900 standards location.
3.
Close the transmission module cover.
4.
Click Calibrate GS-900™ Standards Location.
A progress dialog box shows the calibration in process.
5.
When the GS-900 standards are calibrated, you see the final Instrument
Calibration dialog box prompting you to acquire flat images.
6.
Click Acquire Flat Image.
A progress dialog box shows the acquisition process.