Maintaining the GS-900

Densitometer

Regular cleaning ensures that your GS-900™ calibrated densitometer will continue

to perform optimally. This appendix explains proper care for the GS-900

densitometer.

WARNING! No user-serviceable parts are inside the GS-900 densitometer.

Refer servicing questions to Bio-Rad service personnel.

Cleaning the Outer Surfaces of the GS-900 Densitometer

During normal operation, residue can build up on the sample area.

Use a nonabrasive, lint-free towelette and a 70% ethanol solution to clean the

sample area. Avoid using excessive liquid around the power switch and connector

outlets located on the back side of the unit.

Note: Residual Coomassie stain on the sample area can affect image data. The

glass sample area should be cleaned with a 70% solution of ethanol until no

remaining stain is seen on the towelette.

Resetting the GS-900 Densitometer

If the GS-900 densitometer fails to calibrate, or you notice an uneven appearance to

the background area that has no sample, the Image Lab™ software may need to

relocate the calibration reference location and reset the flat field values.

Note: Ensure that the sample area is clean and free of scratches. Thoroughly

clean the glass sample area of any contamination and residual Coomassie