Default protocols, Setting up and running a default protocol – Bio-Rad Image Lab™ Software User Manual
Page 82

4 | Acquiring Images
82 | GS-900 Calibrated Densitometer with Image Lab Software
Default Protocols
You can image your sample on the GS-900 densitometer by running a default
protocol or a custom protocol. You can set up a default protocol for any gels,
applications, and analyses you are likely to run repeatedly. Then you can access the
protocol with the touch of a button on the front of the densitometer. Like other
protocols, a default protocol can be saved and reused.
First, set up a default protocol for the kind of application you want to run. Then, run
the default protocol, which activates and images the gel with one touch of a button.
Setting Up and Running a Default Protocol
Setting up a default protocol is similar to setting up a new protocol. You can set up a
default protocol
When the instrument is first connected to the software, after initial
calibration
From the Default Protocol button in the toolbar
This procedure assumes that you perform this task using the Default Protocol
button.
To set up a default protocol
1.
Click Default Protocol in the toolbar.
The Default Protocol Setup dialog box appears.
2.
Click Create to create a default protocol.