Bio-Rad Bio-Plex® 200 Systems User Manual

Page 11

Advertising
background image

Table 1. Array Reader Front and Side Panel Features.

Feature Description

Sample arm

The sample arm transports the sample from the 96-well
microtiter plate in the microplate platform to the cuvette.
Upon operation, the carriage drops automatically to the
microtiter plate for sample retrieval.

Sample needle

A stainless-steel sample needle acquires sample from
the 96-well plate in the microplate platform.

Cheminert fitting

Covered by the sample arm cover, this fitting may be
disconnected to allow replacement of the sample needle
if necessary.

Access doors

There are two access doors on the face of the array
reader. The centermost door allows access to the
syringe. The left door provides access to the sheath filter.

Air, waste fluid,

Located on the side of the instrument, these connectors

and sheath fluid

couple directly to the sheath and waste fluid connectors.

connectors

The air connector is green, the sheath connector is blue,
and the waste fluid connector is orange.

The rear panel features of the array reader are shown in Figure 2 and described in Table 2.

Fig. 2. Array reader – rear panel features.

6

www.bio-rad.com/bioplex/

Power connector

Communication
port P1

Communication
port P2

Air filter and
access door

Advertising