Bio-Rad Bio-Plex® 200 Systems User Manual

Page 21

Advertising
background image

16

2.

Insure that the array reader, microplate platform, and HTF unit are turned off.

3.

Insert the Bio-Plex Manager CD-ROM into the CD drive of the computer.

4.

Select Install Bio-Plex Manager.

5.

After installation, remove the installation disk.

6.

Attach the HASP key.

7.

Reconnect the USB communication cable and turn on the array reader, microplate platform,
and HTF unit.

Note: Please refer to the Bio-Plex Manager software manual for detailed installation instructions.

3.7.2 Communication Ports

Bio-Plex Manager will automatically detect the port configuration.

3.8 Installing or Changing the Sample Needle

3.8.1 Installing/Changing the Long Sample Needle

Warning: Turn the power to the array reader off before installing or changing the sample needle.

For use with the microplate platform, a long sample needle (11.7 cm/4.6 in) must be installed. A
spare long needle is shipped with the microplate platform.

1.

Make sure that the power to the array reader is switched off. Make sure the power cord is
unplugged from the outlet.

2.

Remove the light housing directly above the sample arm by grasping and firmly pulling out
(Figure 14). The housing remains attached by a wire. Place the housing on top of the array
reader, out of the way.

Fig. 14. Removing the light housing.

3.

Remove the knurled tubing connector (Cheminert fitting) atop the sample arm by grasping
the sample arm and turning the connector counterclockwise (Figure 15). If the connector is
difficult to remove, push up gently on the sampling needle.