Reconstitute a single vial of diabetes standards, Magpix – Bio-Rad Rat Diabetes Assays User Manual

Page 14

Advertising
background image

12

Assay

Low RP1 (PMT)

High RP1 (PMT)

Human/NHP* diabetes

Mouse diabetes

User validation required*

Rat diabetes

Low RP1 (PMT), broad

High RP1 (PMT), narrow

Cross-panel mixing

range cytokine curve

range cytokine curve

Human diabetes + cytokines

Mouse diabetes + cytokines

User validation required*

Rat diabetes + cytokines

Instrument

RP1 (PMT)

Bio-Plex 100, 200*

Low, high

Bio-Plex 3D*

Standard, enhanced

Bio-Plex MAGPIX*

N/A, use default instrument settings

* Or similar Luminex-based system.

RP1 (PMT) Setting for Standard Curves
The Bio-Plex 200 and 3D systems have two RP1 (PMT or photomultiplier
tube) setting options, while the Bio-Plex

®

MAGPIX

has no PMT and

therefore no PMT setting options. Instead, MAGPIX uses default instrument
settings similar to low PMT on the Bio-Plex 200 (Table 6).

Table 6. Overview of PMT setting options on Bio-Plex systems.

The Bio-Plex Pro diabetes assays were developed on the high PMT
setting using the Bio-Plex 200 system. Protocols using alternative standard
dilution series or low PMT settings should be validated by the end user, for
example when mixing diabetes assays with cytokine assays (Table 7).

Table 7. Settings for optimal sensitivity on the Bio-Plex 200 system*.

* Contact Bio-Rad technical support for the most up-to-date recommendations on PMT

settings and cross-panel multiplexing compatibility.

Reconstitute a Single Vial of Diabetes Standards

This procedure prepares enough material to run each dilution in duplicate.

1. Gently tap the vial containing the lyophilized standard.

2. Add 500 μl of the appropriate standard diluent. Do not use assay
buffer or sample diluent to reconstitute the standards.