Prepare samples, Prepare samples 16, Micro test tubes – Bio-Rad Rat Diabetes Assays User Manual
Page 18: Do not freeze diluted samples
![background image](/manuals/600336/18/background.png)
16
5. Prepare Samples
General guidelines on preparing samples derived from serum, plasma,
and tissue culture supernatant are provided here, including the use of
protease inhibitors with plasma samples.
n
Once thawed, keep samples on ice. Prepare dilutions just prior to the
start of the assay and equilibrate to room temperature before use
n
Prepare sample dilutions in 1.5 or 2 ml polypropylene microcentrifuge
tubes. If a multichannel pipet will be used to load the plate, then aliquot
the required volumes into Titertube
®
micro test tubes.
n
Do not freeze diluted samples
Table 8. Summary of recommended sample diluents and dilution factors.
Sample Type
Diluent
Add BSA
Sample Dilution
Serum and plasma
Sample diluent
None
Fourfold (1:4)
Culture media, with serum
Culture media
None
User optimized
Culture media, serum-free
Culture media
To 0.5% final
User optimized
For Adiponectin Assay
Serum and plasma
Serum-based
None
Human (1:400)
diluent
Mouse (1:1600)
NHP (1:1600)
For Adipsin Assay
Serum and plasma
Serum-based
None
Human (1:400)
diluent
Note: certain sample types may require a different dilution factor.